Intelligence / Healthtech / Device Recalls / Medtronic Minimed, INC.

QualityClass II recall28 Oct 2026

Medtronic Minimed, INC.

28 Oct 2026: Class II recall · CareLink Clinic, REF: MMT-7350. Recorded in the openFDA device enforcement reports.

View the source record

Reported
28 Oct 2026
Recall Class
Class II
Product
CareLink Clinic, REF: MMT-7350
Reason
CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acti
Country
United States
Status
Ongoing
Reference
Z-0203-2027
Location
Northridge, CA