Datex-Ohmeda, INC.

23 Sept 2026: Class II recall · Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000. Recorded in the openFDA device enforcement reports.

View the source record

Reported
23 Sept 2026
Recall Class
Class II
Product
Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000
Reason
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled
Country
United States
Status
Ongoing
Reference
Z-3191-2026
Location
Madison, WI