Intelligence / Healthtech / Device Recalls / Boston Scientific Cork, LTD.

QualityClass II recall28 Oct 2026

Boston Scientific Cork, LTD.

28 Oct 2026: Class II recall · Direxion Torqueable Microcatheter, NGMC/SINGLE/027/STRAIGHT/1RO/130; Material Number: M001195410. Recorded in the openFDA device enforcement reports.

View the source record

Reported
28 Oct 2026
Recall Class
Class II
Product
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/STRAIGHT/1RO/130; Material Number: M001195410
Reason
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Country
Ireland
Status
Ongoing
Reference
Z-0240-2027
Location
Cork, Ireland
What this signal means

A recall puts quality, regulatory, field service and communications under pressure at once.

  1. Class II leads with 660 of 741, then Class I (68).
  2. 235 added in the 30 days to 18 Nov 2026.