Intelligence / Healthtech / Device Recalls / Biomerieux INC

QualityClass II recall30 Sept 2026

Biomerieux INC

30 Sept 2026: Class II recall · VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811. Recorded in the openFDA device enforcement reports.

View the source record

Reported
30 Sept 2026
Recall Class
Class II
Product
VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811
Reason
The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may
Country
United States
Status
Ongoing
Reference
Z-3231-2026
Location
Hazelwood, MO