Beckman Coulter, INC.

21 Oct 2026: Class II recall · IQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325, 700-3326, B48997; IQ200 SELECT, REF: 700-3340, 700-3345, 700-3347, B48998, C10684; IQ200 ELITE, REF: 700-3370. Recorded in the openFDA device enforcement reports.

View the source record

Reported
21 Oct 2026
Recall Class
Class II
Product
IQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325, 700-3326, B48997; IQ200 SELECT, REF: 700-3340, 700-3345, 700-3347, B48998, C10684; IQ200 ELITE, REF: 700-3370
Reason
The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM
Country
United States
Status
Ongoing
Reference
Z-0161-2027
Location
Miami, FL
What this signal means

A recall puts quality, regulatory, field service and communications under pressure at once.

  1. Class II leads with 660 of 741, then Class I (68).
  2. 235 added in the 30 days to 18 Nov 2026.
Reach them with AOS

Find the decision makers at Beckman Coulter, INC.

AOS maps the buying committee, finds verified work emails and drafts outreach that cites this quality signal, for your approval.

Start free

Across AOS Research

Other records for Beckman Coulter, INC. in our registers.

More in Class II

Device Recalls

See all

Source: openFDA device enforcement reports, data to 18 Nov 2026. Device recalls classified by FDA, by report date. Class I is the most serious. AOS shows what the public record says; it does not verify or endorse the company.